| Detectors |
512 Si Array
128 PbS Array
64 PbS Array |
Bandwidth Sampling/
FWHM & Spectral Region |
1.5 nm/3 nm: 350 nm to 1050 nm
6.2 nm/13 nm: 1050nm to 1850nm
8.5 nm/17 nm: 1850nm to 2500 nm |
| Integration Time |
50 ms and up (selectable) |
| FOV |
3° standard, 10° option
25° option with fiber optic |
| Head Size |
320 mm x 335 mm x 320 mm
12.5" x 13.75" x 4.5" |
| Noise Equivalent Radiance |
0.5 s integration time
400 nm: 2.2 x 10-9W · cm-2 ·nm-1· sr-1
700 nm: 1.6 x 10-9W · cm-2 ·nm-1· sr-1
900 nm: 6.2 x 10-9 W · cm-2 ·nm-1 · sr-1
1500 nm: 1.1 x 10-8 W · cm-2 ·nm-1 · sr-1 2200 nm: 1.1 x 10-8 W · cm-2 ·nm-1 · sr-1 |
| Maximum Radiance Levels |
13 ms integration time
700 nm: 1.5 x 10-4W · cm-2 ·nm-1· sr-1 |
Radiometric Calibration Accurancy
(Traceable to NIST) |
400nm : ± 5%
700nm : ± 4%
1000nm : ± 5% 2200nm : ± 7% |
Dark Current Correction
UV-NIR-SWIR |
automatic |
| Humidity |
10 to 90% RH (non-condensing) |